LD 5204 Scale is 2 multi range max=300/600kg d=100/200g and 4 load cells Flintec SB14 454kg. When I put load 100kg on load receptor the instrument LD5204 indicate 100,0 kg, then I live load on load receptor for 2 hours and temperature in room rise for 4°C the instrument LD5204 after that indicate 100,4 kg. Why the indicator is not temperature stable and what do you propose? Answer: The LD5204 weight indicator has excellent drift behaviour that is why it has a min signal per vsi of 0.4μV. Make sure that Par 8.8 =0 in the scale parameter and calibration menu so that ac excitation is selected and the connectors have been soldered correctly. What you observe is the combined effect of temperature effect on zero / sensitivity and CREEP of the load cell and the indicator.
I found out that you have mistake in program because Par 8.7 (load cell amplifier gain adjustment) is 4mV/V and when we change on 2mV/V it nothing happen.